Background There is a shortage of chemical reagents for severe acute respiratory syndrome coronavirus 2 ( SARS-CoV-2 ) diagnosis and a surge of SARS-CoV-2 cases, especially in limited-resource settings. Therefore, the combination of an optimal assay kit is necessary. Methods We compared the ability to screen SARS-CoV-2 among three primer-probe sets in two different master mixes, Invitrogen™ SuperScript™ III One-Step RT-PCR and LightCycler Multiplex RNA Virus Master. Results The assay with TIB-Molbiol, IDT, and Phu Sa sets for LightCycler Multiplex RNA Virus Master or Invitrogen™ SuperScript™ III One-Step RT-PCR showed positive results from a single reaction of triplicate in the three days of 4.8 copies per reaction. R squared and amplification efficiency were 0.97 and ranged from 107 to 108%, respectively. Conclusions Our findings indicated that TIB-Molbiol, IDT, and Phu Sa primer-probe sets could be beneficial for the laboratory screening of SARS-CoV-2 by RT-qPCR assay of E gene. There is a need to consider the combination of these reagent sets as a new strategy to increase the testing capacity of screening programs for COVID-19.
【초록키워드】 COVID-19, SARS-CoV-2, coronavirus, Diagnosis, RT-PCR, Laboratory, RNA, amplification, RT-qPCR, Combination, Efficiency, acute respiratory syndrome, E gene, positive result, reagent, Result, indicated, ranged, IDT, primer-probe set, 【제목키워드】 SARS-CoV-2, coronavirus 2, Screening, respiratory, Mixe,