[[[ Aim: ]]] The aim was to assess the flagging performance of Sysmex XN-10 haematology analyser for malaria detection through the parasitic red blood cell (‘pRBC’) alarm. [[[ Methods: ]]] We retrospectively studied 584 blood samples performed on the Sysmex XN-10 analyser that were tested for malaria. Sensitivity, specificity, positive and negative predictive values, and prevalence were established for the pRBC alarm. [[[ Results: ]]] Sensitivity, specificity, and positive and negative predictive values for the pRBC flag were 7.8%, 100%, 100% and 87.7%, respectively. The prevalence of pRBC flag of 0.026% in the overall population was significantly different from the prevalence of 1.027% in the population tested for malaria. [[[ Conclusions: ]]] Considering the excellent specificity and the low prevalence of the flag in the overall population, we suggest, in case of the presence of pRBC flag, the implementation of a rapid review of the blood smear looking for Plasmodium , mostly if the patient had fever and had not been tested for malaria.
Flagging performance of Sysmex XN-10 haematology analyser for malaria detection