The rapid identification and classification of pathogenic microorganisms, including Salmonella enterica, is important for the surveillance and prevention of foodborne diseases. Matrix-assisted laser desorption\ionization time-of-flight mass spectrometry (MALDI-TOFMS) has been shown to be an effective tool for the rapid identification of microorganisms. In a previous report, a mass database consisting of 12 biomarker proteins, S8, L15, L17, L21, L25, S7, superoxide dismutase (SodA), peptidylprolyl cis-trans isomerase C, Gns, YibT, YaiA, and YciF, was introduced for the serotyping of S. enterica via MALDI-MS (Applied Microbiology and Biotechnology, 2017, 101, 8557-8569). However, the reproducibility of peak detection of biomarkers such as SodA at m\z 23 000 was poor. We report here an optimized MALDI-MS method for detecting these biomarkers with high sensitivity and reproducibility. The issue was solved by controlling the bacterial concentration at 1 × 10 to 1 × 10^{2} MFU (3 × 10^{6} to 3 × 10^{7} CFU\μL, as calculated from the MFU), using the colony suspension supernatant obtained by centrifugation, and using matrix additives such as methylenediphosphonic acid and N-decyl-β-D-maltopyranoside. We propose that the method including the above steps is one of the best for detecting biomarkers with high sensitivity and reproducibility.
【저자키워드】 Biomarkers, Salmonella, serotype, MALDI-MS, proteotyping,