Abstract Objective To evaluate the performance of new lateral flow immunoassays (LFIAs) suitable for use in a national coronavirus disease 2019 (covid-19) seroprevalence programme (real time assessment of community transmission 2—React 2). Design Diagnostic accuracy study. Setting Laboratory analyses were performed in the United Kingdom at Imperial College, London and university facilities in London. Research clinics for finger prick sampling were run in two affiliated NHS trusts. Participants Sensitivity analyses were performed on sera stored from 320 previous participants in the React 2 programme with confirmed previous severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection. Specificity analyses were performed on 1000 prepandemic serum samples. 100 new participants with confirmed previous SARS-CoV-2 infection attended study clinics for finger prick testing. Interventions Laboratory sensitivity and specificity analyses were performed for seven LFIAs on a minimum of 200 serum samples from participants with confirmed SARS-CoV-2 infection and 500 prepandemic serum samples, respectively. Three LFIAs were found to have a laboratory sensitivity superior to the finger prick sensitivity of the LFIA currently used in React 2 seroprevalence studies (84%). These LFIAs were then further evaluated through finger prick testing on participants with confirmed previous SARS-CoV-2 infection: two LFIAs (Surescreen, Panbio) were evaluated in clinics in June-July 2020 and the third LFIA (AbC-19) in September 2020. A spike protein enzyme linked immunoassay and hybrid double antigen binding assay were used as laboratory reference standards. Main outcome measures The accuracy of LFIAs in detecting immunoglobulin G (IgG) antibodies to SARS-CoV-2 compared with two reference standards. Results The sensitivity and specificity of seven new LFIAs that were analysed using sera varied from 69% to 100%, and from 98.6% to 100%, respectively (compared with the two reference standards). Sensitivity on finger prick testing was 77% (95% confidence interval 61.4% to 88.2%) for Panbio, 86% (72.7% to 94.8%) for Surescreen, and 69% (53.8% to 81.3%) for AbC-19 compared with the reference standards. Sensitivity for sera from matched clinical samples performed on AbC-19 was significantly higher with serum than finger prick at 92% (80.0% to 97.7%, P=0.01). Antibody titres varied considerably among cohorts. The numbers of positive samples identified by finger prick in the lowest antibody titre quarter varied among LFIAs. Conclusions One new LFIA was identified with clinical performance suitable for potential inclusion in seroprevalence studies. However, none of the LFIAs tested had clearly superior performance to the LFIA currently used in React 2 seroprevalence surveys, and none showed sufficient sensitivity and specificity to be considered for routine clinical use.
【초록키워드】 coronavirus disease, SARS-CoV-2, IgG, Coronavirus disease 2019, coronavirus, antibody, SARS-COV-2 infection, Infection, Spike protein, clinics, Laboratory, immunoassay, lateral flow immunoassay, serum, sensitivity, Immunoglobulin G, Diagnostic accuracy, Immunoglobulin, Accuracy, Sensitivity and specificity, sera, community transmission, LFIA, respiratory, United Kingdom, London, Antibody titre, Analysis, real time, Reference standards, Finger prick, Clinical use, NHS, acute respiratory syndrome, antigen binding, acute respiratory syndrome coronavirus, serum samples, acute respiratory syndrome coronavirus 2, 95% confidence interval, antibody titres, enzyme, laboratory analyses, Sensitivity analyses, finger, outcome measure, cohorts, participant, titre, minimum, National, college, PanBio, LFIAs, serum sample, Surescreen, positive sample, Inclusion, objective, setting, Seven, lowest, clinical sample, Result, tested, performed, evaluate, evaluated, analysed, were used, significantly higher, antigen binding assay, Laboratory analysis, 【제목키워드】 SARS-CoV-2, diagnostic, lateral flow, Accuracy, National, seroprevalence survey,