SARS-CoV-2 real-time reverse-transcription PCR (rRT-PCR) is the most effective testing system currently available to counter COVID-19 epidemics when potent treatments and vaccines are unavailable. Therefore, four SARS-CoV-2 rRT-PCR kits have been approved by the emergency-use-authorization (EUA) without clinical validation in Korea until March 15, 2020. This study evaluated the analytical and clinical performance of these kits. Allplex 2019-nCoV Real-time PCR (Seegene, Seoul, Korea), PowerChek 2019-nCoV (KogeneBiotech, Seoul), Real-Q 2019-nCoV Real-Time Detection (BioSewoom, Seoul), and StandardM nCoV Detection (SD BIOSENSOR, Osong, Korea) were evaluated. The limit of detection (LODs) of Allplex, PowerChek, and Real-Q was determined by testing the transcribed RNA of SARS-CoV-2 E and the RNA of SARS-CoV Frankfurt1. A total of 27 consecutive samples comprising 13 sputum, 12 nasopharyngeal swab (NPS), 1 urine and 1 stool sample were collected from 2 COVID-19 patients for sensitivity analysis. Precision was assessed via daily tests of positive and negative controls in each kit for 5 d. Reproducibility was examined by repeating 21 samples and 10-fold dilutions of 14 samples in pairs using Allplex. Specificity was evaluated with 24 other respiratory virus-positive samples. LOD of Allplex, PowerChek, and Real-Q were 153.9, 84.1, and 80.6 copies/mL, respectively. The degrees of association between Cts and log viral concentrations by Allplex and PowerChek was expressed as y = −3.319 log (x) + 42.039 (R = 0.96) and y = −3.392 log(x) + 43.113 (R = 0.98), respectively. One or more of the 4 kits detected 20 out of 27 clinical samples positive. Of the 20 positive samples, the detection rates of positives for Allplex, PowerChek, Real-Q, and StandardM were 90.0, 82.3, 75.0, and 100.0%, respectively, but those of PowerChek and Real-Q would be 100% if out-of-cutoff Cts were counted as positives. Precision was 100%. Interpretation of Allplex results was reproducible when Ct of E ≤33. All 4 kits showed no cross-reactivity with other respiratory viruses. Performance of the 4 kits indicated the suitability of these for diagnosis and follow-up testing of COVID-19. Laboratory doctors who initially implement these EUA kits must be able to interpret quality control parameters.
【저자키워드】 severe acute respiratory syndrome coronavirus 2, polymerase chain reaction, reverse transcription, real time, Coronavirus infectious disease 2019, emergency-use-authorization, 【초록키워드】 COVID-19, Treatment, SARS-CoV-2, Vaccine, SARS-CoV, 2019-nCoV, Epidemics, Diagnosis, detection, sputum, EUA, RNA, cross-reactivity, Nasopharyngeal swab, Stool, rRT-PCR, PCR, limit of detection, Urine, sensitivity analysis, Follow-up, Quality control, performance, detection rate, parameters, association, COVID-19 patient, Dilution, Viral concentration, nCoV, positive, Real-time reverse-transcription PCR, NPs, effective, consecutive sample, negative control, log, clinical sample, collected, examined, indicated, evaluated, approved, was determined, expressed, transcribed, 【제목키워드】 SARS-CoV-2, reaction, chain,