Since autumn 2020, rapid antigen tests (RATs) have been implemented in several countries as an important pillar of the national testing strategy to rapidly screen for infections on site during the SARS-CoV-2 pandemic. The current surge in infection rates around the globe is driven by the variant of concern (VoC) omicron (B.1.1.529). Here, we evaluated the performance of nine SARS-CoV-2 RATs in a single-centre laboratory study. We examined a total of 115 SARS-CoV-2 PCR-negative and 166 SARS-CoV-2 PCR-positive respiratory swab samples (101 omicron, 65 delta (B.1.617.2)) collected from October 2021 until January 2022 as well as cell culture-expanded clinical isolates of both VoCs. In an assessment of the analytical sensitivity in clinical specimen, the 50% limit of detection (LoD50) ranged from 1.77 × 10 6 to 7.03 × 10 7 RNA copies subjected to the RAT for omicron compared to 1.32 × 10 5 to 2.05 × 10 6 for delta. To score positive in these point-of-care tests, up to 10-fold (LoD50) or 101-fold (LoD95) higher virus loads were required for omicron- compared to delta-containing samples. The rates of true positive test results for omicron samples in the highest virus load category (Ct values < 25) ranged between 31.4 and 77.8%, while they dropped to 0–8.3% for samples with intermediate Ct values (25–30). Of note, testing of expanded virus stocks suggested a comparable RAT sensitivity of both VoCs, questioning the predictive value of this type of in vitro-studies for clinical performance. Given their importance for national test strategies in the current omicron wave, awareness must be increased for the reduced detection rate of omicron infections by RATs and a short list of suitable RATs that fulfill the minimal requirements of performance should be rapidly disclosed. Supplementary Information The online version contains supplementary material available at 10.1007/s00430-022-00730-z.
【저자키워드】 VoC, diagnostic test, nucleocapsid protein, lateral flow, sensitivity, specificity, SARS-CoV-2 rapid antigen test, 【초록키워드】 SARS-CoV-2, pandemic, SARS-CoV-2 pandemic, variant, Infection, B.1.617.2, omicron, Predictive value, Laboratory, Rapid antigen tests, rapid antigen test, RNA copies, Cell culture, limit of detection, Awareness, B.1.1.529, infection rate, respiratory, virus load, detection rate, analytical sensitivity, RAT, Ct values, Predictive, test results, supplementary material, infection rates, PCR-positive, RATs, specimen, positive test result, National, positive, point-of-care tests, PCR-negative, isolate, country, Cell, swab sample, highest, collected, examined, globe, evaluated, required, nine, reduced, RNA copy, suggested, comparable, driven by, virus stock, ranged, dropped, the SARS-CoV-2, 【제목키워드】 SARS-CoV-2, rapid antigen test, Impaired,